Prepare Patient for Defibrillation
Initial defibrillation:
Apply OneStep CPR 1 A/A Multifunctional pads. The pads should be connected to the Zoll and the package unopened. Before starting, confirm that the pads and machine have passed their self-test and are ready to open and use.
If CPR is in progress, apply the gel pads in the Anterior - Anterior position.
Anterior-Anterior Placement
Apply Pad #1 first to the left chest position. Position below the pectoral muscle as shown on pad diagram. For female patients, lift the breast and position on top of the pad once it has been applied.
Pad #2 is attached to a CPR sensor. Position the CPR sensor so that the red vertical line is mid-sternum and the blue horizontal line is aligned with the nipple. Press the right pad into position as shown on the pad diagram.
|
|
|
OneStep Multifunctional CPR A/A Pads can be used for either Anterior-Anterior or Anterior-Posterior. Pads remain connected to Zoll cable and remain unopened until use. |
Apply apex pad first. For female patients, lift the breast and position the pad under the breast. |
Position cross bars of CPR sensor to align with the mid sternum and nipple line. The right chest pad is attached to the CPR sensor for correct positioning.
If Anterior-Posterior placement is desired, separate the right chest pad from the sensor and apply to the patients left back. Note: the posterior pad should be placed first.
|
Anterior-Posterior Placement
The OneStep CPR 1 A/A Multifunctional pads can be used for anterior - posterior placement. Separate the right chest pad from the CPR sensor. Apply the right chest pad to the left posterior chest first (to prevent wrinkling of the anterior pads. Apply the apex pad as shown above. Apply the apex pad as shown above.
|
|
|
Identify the checkmark in the window display. This confirms that the pads and machine are functioning prior to use. |
The purple display identifies the quality and depth of CPR. |
End Tidal CO2confirms adequacy of CPR and endotracheal tube placement. |
ECG Electrodes
Apply the Zoll ECG electrodes. The One Step Multifunctional pads can monitor, shock and pace, but they cannot read the ECG at the same time that they pace or cardiovert. The ECG must be connected to the Zoll to provide sending and rhythm feedback.
Paddles are no longer available on the Crash Carts.
Prepare Patient:
- Side rails.
- Bed and patient flat.
- Chest exposed. When time permits, hair should be clipped (not shaved).
- Confirm gel pads in correct position and pressed onto skin.
- Ensure paddles have no contact with metal jewelry, pharmacological patches (e.g., fentanyl, nitroglycerin) or transparent dressings..
Position for Pacemaker/Defibrillator Pads:
Safety:
OneStep CPR 1 A/A Multifunctional pads are attached and ready to open and apply on all Crash Carts. When the unopened package is connected to the Zoll R, they perform ongoing self-tests to ensure functionality. The package must remain unopened for the self-test to work. Confirm that the pads and machine are functional prior to use by confirming the check mark is visible in the display window.
The CPR A/A pads can be applied either in the Anterior-Posterior position or Anterior-Anterior, however, during CPR they should be applied Anterior-Anterior for speed.
When pads are applied Anterior-Posterior, apply the posterior pad first to avoid wrinkling of the anterior pad.
Do no place pads over ECG electrodes. Arcing will occur if shock is delivered too close to electrodes.
|